Instructions for the
Tritech Research MICROMAGNIPULATOR™
Microinjection Needle
Positioning System
WARNINGS!!!!
THE MICROMAGNIPULATOR™ BASE
IS A CUBE OF SOLID STEEL WEIGHING OVER 22 POUNDS (10 KG); IT IS SMALL, HAS
SHARP EDGES AND CAN BE DIFFICULT TO MOVE.
TAKE CARE NOT TO DROP IT OR STRAIN YOUR HANDS, ARMS, OR BACK WHILE
LIFTING IT. THE POLISHED SURFACE ON THE
BLOCK MUST BE GREASED BEFORE ASSEMBLY, OR THE SURFACE WILL BE SCRATCHED. NEVER TRY TO MOVE THE BASE BY HOLDING ONTO
THE MAGNETIC NEEDLE HOLDER ASSEMBLY; IT
MAY NOT HOLD THE WEIGHT, AND BECAUSE THE INTERFACE IS GREASED, THE BASE WILL
ALMOST CERTAINLY SLIDE OFF!
SETTING UP:
As stated in the WARNINGS above, the MICROMAGNIPULATOR™ base is heavy, so take great care when removing it
from the box and unpacking it to avoid injury.
Also, be careful not to scratch the surface on top. Many possible configurations are possible,
but to start with, position the base to the RIGHT side of the microinjection
microscope, with the Tritech Research sticker facing the user.
Install the three
X - Y positioning bolts a few turns into the three nuts on the base.
Unpack the
magnetic needle holder / Z - control assembly, taking care not to displace the
glide disk on the bottom. So as not to
be damaged, the top of the base and bottom of the magnetic holder assembly MUST be greased thoroughly before they
are put together. A highly viscous
grease, such as silicone-based vacuum grease (Dow Corning, Beckman, etc.) is
preferable and will give smoothest operation and longest life. Petroleum jelly is an acceptable, but less
desirable alternative.
Make sure the
magnet control knob is set to OFF, and carefully place the magnetic holder
assembly on the base between the three X - Y positioning bolts, with the cotton
swab facing the microscope. Turn the
magnet control knob toward ON. The
cotton swab is there to show you where the microinjection needle holder should
be installed. A “straight arm” needle
holder, such as the solid brass needle holder available with the Tritech
Research microINJECTOR™ System is ideal.
Adjust the
position of the MICROMAGNIPULATOR ™ and needle holder such that
the needle holder is slightly above the microscope stage, coming down at a
shallow angle. If extra height is
necessary, the MICROMAGNIPULATOR™ can be placed on a block of marble, or other
heavy, stable object.
USE:
Tightly install the needle holder,
containing the microinjection pipette, between the holding washers, using the
thumb screw. Position the end of the
pipette in the general vicinity of the microscopic viewing area using the pole
adjustments and by where you install the needle holder.
Practice sliding
the magnetic holder assembly on its base by hand, adjusting the position of the
magnetic control knob until the magnetic holder assembly moves slowly,
smoothly, and does not stick in place.
To move directly by hand, firmly grasp the magnetic holding cube and
slowly apply pressure. Twist the cube slightly
if it is stuck initially.
Adjust the
pipette’s X - Y position by sliding the magnetic holder assembly on the
base. This can be accomplished either
by sliding the magnetic holder assembly directly by hand, or by use of the X -
Y positioning bolts.
Adjust the
pipette’s Z position, and angle, with the Z adjustment knob located on the
horizontal pole. There are a variety of
swivel points, so feel free to completely reconfigure the poles and swivels to
best suit your positioning / micromanipulating needs.
TRITECH RESEARCH, INC. 2961 Veteran Avenue, Los Angeles, CA 90064 (310) 446-4460 (800) 7-333-TRI